The comparative stability of screw versus plate versus screw and plate coronoid fixation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21276887.
- Also identified by DOI 10.1016/j.jhsa.2010.10.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the biomechanical characteristics of screw versus plate versus both screw and plate fixation for large, type 3 O'Driscoll coronoid fractures. Synthetic ulnas had 70% of their coronoids cut. Fixation was performed with either a cannulated screw, a plate, or both a screw and a plate. Energy to failure, force at failure, first cycle stiffness, and stiffness at failure were measured on a servohydraulic testing machine under cyclic posterior axial loading. The combination of a plate and screw had significantly greater energy to failure (83 Nm), force required to cause failure (634 N), and stiffness at failure (387 N/mm) compared to either an isolated plate (38 Nm, 474 N, 237 N/mm, respectively) or a screw (10 Nm, 279 N, 149 N/mm, respectively). For energy to failure and force required to cause failure, the plate group significantly outperformed the screw group. There was no significant difference in stiffness at the time of failure between the plate and screw groups. For type 3 O'Driscoll coronoid fractures or nonunions when both a screw and a plate can be placed, the combination of these 2 fixation devices appears to produce significantly greater biomechanical stability than either fixation device alone.
Medical subject headings
- Bone Plates
- Bone Screws
- Fracture Fixation, Internal