Hydrodynamic optical alignment for microflow cytometry.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21279198.
- Also identified by DOI 10.1039/c0lc00500b and PMC identifier 3684690.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A microfabricated flow cytometer has been developed that is capable of detecting nearly all of the microparticles in an aqueous suspension. Current design allows for integrated coupling between an optical fiber-based detection system and the particle stream via hydrodynamic focusing. By adjusting the relative flow-rates at the auxiliary inputs of the focusing manifold, the particle stream can be steered out-of-plane relative to the illuminating laser, and similarly the particle stream can be squeezed or expanded. The microfabricated device was constructed in polydimethylsiloxane with cross-sectional microfluidic dimensions of 125 µm×125 µm. Using the present device and method, fluorescent microparticles in aqueous solution were counted at an absolute counting efficiency of 91±4%. The coefficient of variation of the fluorescence pulse-heights for far-red fluorescent microparticles was 15%. The device exhibited a linear response to fluorescence intensity calibration microparticles as shown by comparison with a commercial cytometer instrument.
Medical subject headings
- Flow Cytometry