A microfabricated electrical differential counter for the selective enumeration of CD4+ T lymphocytes.

Watkins, Nicholas N; Sridhar, Supriya; Cheng, Xuanhong; Chen, Grace D; Toner, Mehmet; Rodriguez, William; Bashir, Rashid · Lab Chip · 2011

basic_science · Level V

Where this comes from

Abstract

We have developed a microfabricated biochip that enumerates CD4+ T lymphocytes from leukocyte populations obtained from human blood samples using electrical impedance sensing and immunoaffinity chromatography. T cell counts are found by obtaining the difference between the number of leukocytes before and after depleting CD4+ T cells with immobilized antibodies in a capture chamber. This differential counting technique has been validated to analyze physiological concentrations of leukocytes with an accuracy of ∼9 cells per µL by passivating the capture chamber with bovine serum albumin. In addition, the counter provided T cell counts which correlated closely with an optical control (R(2) = 0.997) for CD4 cell concentrations ranging from approximately 100 to 700 cells per µL. We believe that this approach can be a promising method for bringing quantitative HIV/AIDS diagnostics to resource-poor regions in the world.

Medical subject headings