Diagnostic accuracy of point-of-care testing for diabetic ketoacidosis at emergency-department triage: {beta}-hydroxybutyrate versus the urine dipstick.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21307381.
- Also identified by DOI 10.2337/dc10-1844 and PMC identifier 3064039.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In the emergency department, hyperglycemic patients are screened for diabetic ketoacidosis (DKA) via a urine dipstick. In this prospective study, we compared the test characteristics of point-of-care β-hydroxybutyrate (β-OHB) analysis with the urine dipstick. Emergency-department patients with blood glucose ≥250 mg/dL had urine dipstick, chemistry panel, venous blood gas, and capillary β-OHB measurements. DKA was diagnosed according to American Diabetes Association criteria. Of 516 hyperglycemic subjects, 54 had DKA. The urine dipstick had a sensitivity of 98.1% (95% CI 90.1-100), a specificity of 35.1% (30.7-39.6), a positive predictive value of 15% (11.5-19.2), and a negative predictive value of 99.4% (96.6-100) for DKA. Using the manufacturer-suggested cutoff of >1.5 mmol/L, β-OHB had a sensitivity of 98.1% (90.1-100), a specificity of 78.6% (74.5-82.2), a positive predictive value of 34.9% (27.3-43), and a negative predictive value of 99.7% (98.5-100) for DKA. Point-of-care β-OHB and the urine dipstick are equally sensitive for detecting DKA (98.1%). However, β-OHB is more specific (78.6 vs. 35.1%), offering the potential to significantly reduce unnecessary DKA work-ups among hyperglycemic patients in the emergency department.
Medical subject headings
- 3-Hydroxybutyric Acid
- Diabetic Ketoacidosis
- Emergency Service, Hospital
- Point-of-Care Systems
- Triage