Implant survival analysis and failure modes of the X-Stop interspinous distraction device.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 21311406.
- Also identified by DOI 10.1097/BRS.0b013e31820b86e1.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Retrospective cohort study. To review the clinical outcome, implant survivorship and reasons for failure after X-Stop implantation. Conflicting data exist concerning clinical outcome of the X-Stop interspinous spacer and little information is being published about implant survivorship and the need for revision surgery. This retrospective review evaluated 46 patients who underwent implantation of the X-Stop interspinous spacer for the treatment of neurogenic claudication. After a mean follow-up of 40 months, pain levels, 36-Item Short Form Health Survey (SF-36), Oswestry Disability Index (ODI), and the need for subsequent surgery were assessed and a Kaplan-Meier survivorship analysis was performed. Within the follow-up period, the revision rate was found to be 30.4%. Lack of improvement at 6-week follow-up correlated well with subsequent revision surgery, which predominantly took place within 12 months after the index surgery. In patients who did not need to undergo revision surgery, clinical outcome parameters improved significantly. Kaplan-Meier survivorship analysis predicted an implant survival probability of 0.68 at 48 months postoperatively. Clinical outcome after X-Stop implantation might be considerably less favorable than when it was being published previously. Patient selection might be a reason for early revision surgery. More criteria for better X-Stop indications might be needed.
Medical subject headings
- Intermittent Claudication
- Lumbar Vertebrae
- Neuralgia
- Prostheses and Implants
- Spinal Stenosis
Anatomy
- lumbar spine