Straight tapered titanium stem with alumina bearing in cementless primary total hip arthroplasty: a minimum 5-year follow-up.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 21316911.
- Also identified by DOI 10.1016/j.arth.2010.12.015.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to assess the midterm results of primary cementless total hip arthroplasty using a tapered stem and alumina bearing couple in active patients. After a minimum of 5 years of follow-up, 78 arthroplasties in 72 patients were reviewed retrospectively. The mean Harris hip score was 94 points, and 2 hips had thigh pain. All components radiographically demonstrated stable fixation by bone ingrowth and mild stress shielding of the proximal femur were noted in 14% of hips. There was no significant osteolysis or aseptic loosening. There was a ceramic head fracture in 1 hip and audible sounds in 2 hips. The results of total hip arthroplasty with a straight, tapered, proximally porous-coated stem and alumina-on-alumina bearing were encouraging for active patients.
Medical subject headings
- Aluminum Oxide
- Arthroplasty, Replacement, Hip
- Femur
- Hip Prosthesis
- Osteonecrosis
- Prosthesis Design
- Titanium
Anatomy
- femur
- hip