Micromotion of plasma spray versus grit-blasted radial head prosthetic stem surfaces.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21324417.
- Also identified by DOI 10.1016/j.jse.2010.11.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Initial stability of a textured surface prosthetic stem is necessary for bone in-growth. Surfaces currently used for radial head prostheses include titanium plasma spray and grit-blasted titanium. Plasma spray radial head prosthetic stems are less dependent than grit-blasted stems on a tight press fit. Good initial press-fit stability, with acceptable micro-motion, can be achieved with a greater range of stem sizes using a plasma spray than grit-blasted surface. Paired cadaveric radii were implanted with plasma spray or grit-blasted radial head prosthetic stems. Micromotion at the stem tip was measured under circumstances simulating eccentric loads. Micromotion in the plasma spray (PS) stems (49 ± 37) μm was not better than that in the grit-blasted (GB) stems (28 ± 10) μm (P = .13). Micromotion of less than 100 μm was measured in all 12 GB stems that were maximum or 1 mm less than maximum size, versus 5/6, and 4/6 PS stems, respectively. Micromotion in plasma spray prosthetic radial head stems was not better than that seen in grit-blasted stems, contrary to our initial hypothesis. Grit-blasted prosthetic radial head stems confer initial press-fit stability that is as good as, or slightly better than, corresponding plasma spray stems. Acceptable amounts of micromotion can be achieved with 2 grit-blasted stem sizes and probably with 2 plasma spray stem sizes.
Medical subject headings
- Arthroplasty, Replacement, Elbow
- Elbow Prosthesis
- Osseointegration
- Radius
- Range of Motion, Articular
Anatomy
- radius
- elbow