Shear-mediated platelet adhesion analysis in less than 100 μl of blood: toward a POC platelet diagnostic.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21342809.
- Also identified by DOI 10.1109/TBME.2010.2090659.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a microfluidic chip-based hydrodynamic focusing approach that minimizes sample volume for the analysis of cell-surface interactions under controlled fluid-shear conditions. Assays of statistically meaningful numbers of translocating platelets interacting with immobilized von Willebrand factor at arterial shear rates (∼1500 s(-1)) are demonstrated. By controlling spatial disposition and relative flow rates of two contacting fluid streams, e.g., sample (blood) and aqueous buffer, on-chip hydrodynamic focusing guides the cell-containing stream across the protein surface as a thin fluid layer, consuming ∼50 μL of undiluted whole blood for a 2-min platelet assay. Control of wall shear stress is independent of sample consumption for a given flow time. The device design implements a mass-manufacturable fabrication approach. Fluorescent labeling of cells enables readout using standard microscopy tools. Customized image-analysis software rapidly quantifies cellular surface coverage and aggregate size distributions as a function of time during blood-flow analyses, facilitating assessment of drug treatment efficacy or diagnosis of disease state.
Medical subject headings
- Blood Platelets
- Microfluidic Analytical Techniques
- Platelet Adhesiveness
- Point-of-Care Systems