Evaluation of the new Statscan radiography device for ventriculoperitoneal shunt assessment.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21343476.
- Also identified by DOI 10.2214/AJR.10.4939.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We assessed the diagnostic accuracy and time effectiveness of the Statscan low-dose full-body digital x-ray machine in the evaluation of ventriculoperitoneal (VP) shunt dysfunction. Twenty-five consecutive adults scheduled for a radiographic shunt series with the Statscan were prospectively selected (group 1). Another 21 consecutive patients who underwent conventional x-ray scanning served as controls (group 2) for blinded analysis. Image quality was sufficient for diagnosis in both groups (95.6% vs 95.1%). The quality of VP shunt delineation and contrast of anatomic structures was good or optimal in both groups (p > 0.05). The scan time was significantly lower for group 1 (p < 0.001). All images for group 2 exhibited overlapping of anatomic structures (25% of total coverage), whereas no technically determined overlapping occurred in group 1. The Statscan digital x-ray machine allows accurate assessment of VP shunt continuity at lower radiation dosages and in less time than conventional x-ray techniques.
Medical subject headings
- Radiographic Image Enhancement
- Ventriculoperitoneal Shunt
- Whole Body Imaging