Effect of a new pelvic stabilizer (T-POD®) on reduction of pelvic volume and haemodynamic stability in unstable pelvic fractures.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21374905.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Pelvic fractures, often the result of high energy blunt trauma, are associated with severe morbidity and mortality. A new pelvic stabilizer (T-POD®) provides secure and effective simultaneous circumferential compression of the pelvis. In this study we describe 15 patients with a prehospital untreated unstable pelvic fracture with signs of hypovolaemic shock with the T-POD®. Before and 2 min after applying the T-POD®, heart rate and blood pressure were measured. An X-ray before and directly after applying the T-POD® was made to measure the effect on reduction in symphyseal diastasis. Application of the T-POD® reduced the symphyseal diastasis with 60% (p = 0.01). The mean arterial pressure (MAP) increased significant from 65.3 to 81.2 mm Hg (p = 0.03) and the heart rate declined from 107 beats per minute to 94 (p = 0.02). Out of ten patients in whom the circulatory response before and after the T-POD® was recorded, seven were good responders, one had a transient response and two responded poor. In the acute setting, the T-POD® device has a clear compressive effect on the pelvic volume in unstable pelvic fractures. The T-POD® is therefore an effective and easy to use device in (temporarily) stabilizing the pelvic ring in haemodynamically unstable patients.
Medical subject headings
- Fracture Fixation
- Fractures, Bone
- Fractures, Compression
- Orthotic Devices
- Pelvic Bones