Gender analysis of the pivotal results of the Medtronic Talent Thoracic Stent Graft System (VALOR) trial.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21397440.
- Also identified by DOI 10.1016/j.jvs.2010.12.064.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study evaluated the differences between male and female patients undergoing thoracic endovascular aneurysm repair (TEVAR) in a pivotal Food and Drug Administration (FDA)-approved trial. The Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) study was a prospective, nonrandomized, multicenter, pivotal trial conducted in the United States. Patients were enrolled between December 2003 and June 2005. Follow-up was conducted at 30 and 365 days. VALOR enrolled 115 men (58.9%; 69.3 ± 11.7 years old), and 80 women (41.1%; 71.6 ± 10.1 years old). Iliac conduits were used more often in women, who had smaller diameter external iliac arteries, than in men (38.8% vs 8.8%, P < .001). Women required more blood transfusions and had a longer hospital length of stay. At 30 days, more major adverse events occurred in women than in men (52.5% vs 33.0%, P = .008), with more vascular access-related and respiratory complications. No gender-based differences were seen in all-cause mortality or in aneurysm-related death. The composite end point of 365-day "successful aneurysm treatment," defined as no aneurysm growth >5 mm at the 365-day follow-up visit compared with the 30-day follow-up visit and absence of any type I endoleak requiring a secondary procedure, favored women over men (98.2% vs 82.4%, P = .004). TEVAR with the Talent device provided similar rates of 365-day mortality and morbidity for men and women. Although female patients had higher rates of periprocedural complications, they also more often had successful aneurysm treatment at the 1-year follow-up.
Medical subject headings
- Aortic Aneurysm, Thoracic
- Blood Vessel Prosthesis
- Blood Vessel Prosthesis Implantation
- Endovascular Procedures
- Stents