Statistical estimation of EIT electrode contact impedance using magic Toeplitz matrix.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21402505.
- Also identified by DOI 10.1109/TBME.2011.2125790 and PMC identifier 3233639.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The goal of the paper is to propose a fast and reliable method of simultaneous estimation of conductivity and electrode contact impedances for a homogeneous 2D disk. Magic Toeplitz matrix as the Neumann-to-Dirichlet map with finite width electrodes plays the central role in our linear model, called the gapZ model. This model enables testing of various hypotheses using the F-test, such as the uniformity of electrode impedances and their statistical significance. The gapZ model is compared with the finite element approximation, and illustrated and validated with a phantom tank experiment filled with saline. Further this model was illustrated with the patient breast EIT data to identify bad contact electrodes.
Medical subject headings
- Algorithms
- Breast
- Dielectric Spectroscopy
- Electrodes
- Equipment Failure Analysis
- Models, Biological