Exposure from the large C-arm versus the mini C-arm using hand/wrist and elbow phantoms.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21414730.
- Also identified by DOI 10.1016/j.jhsa.2011.01.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study tests the conventional wisdom that using fluoroscopy under identical geometrical conditions results in less radiation when using the mini C-arm relative to the large C-arm. We evaluated the radiation dose for both direct exposure and scatter 2.54 cm outside the intensifier. We used 3 mini and 3 large C-arms in a vertical orientation with the image intensifier below the specimen and the source above. We used 2 specimens: a cadaver hand/wrist and a cadaver elbow. Specimens were tested both directly on the intensifier and on a hand table placed on the intensifier. For the same setup, use of the mini C-arm resulted in direct patient radiation exposure greater than the exposure delivered by the large C-arm. Specifically, exposure using the mini C-arm was 53% to 70% greater than that using the large C-arm. In addition, use of the hand table resulted in exposure 80% to 94% greater compared with placing the specimen directly on the intensifier. In all cases, scatter at 2.54 cm from the intensifier resulted in an average exposure of 1.5% (SD, 0.24%) of the direct beam. Tube current, and therefore machine radiation output, was approximately 13 to 14 times greater for the large C-arm. Direct radiation exposure to the patient and scatter to the surgeon are minimized when the C-arm is positioned with the intensifier below and the extremity is placed directly on the intensifier. Under identical geometrical conditions with the intensifier below the specimen, the large C-arm with its greater source to image intensifier distance is associated with less radiation exposure than the mini C-arm.
Medical subject headings
- Phantoms, Imaging
- Radiation Dosage
- Radiation Protection
Anatomy
- hand
- wrist
- elbow