Ultrasound measurement of aortic diameter in a national screening programme.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21439859.
- Also identified by DOI 10.1016/j.ejvs.2011.02.030.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Currently there is no universally accepted standard for ultrasound measurement of abdominal aortic aneurysm (AAA). The aim was to investigate the reliability and reproducibility of inner to inner (ITI) versus outer to outer (OTO) ultrasound measurement of AAA diameter. A prospective study design was used to collect 60 random images of aorta (1.4-7.1 cm). Inner and outer wall diameter measurements were then performed by 13 qualified AAA screening technicians and 11 vascular sonographers. The mean (range) diameter for all 60 aortas by ITI was 3.91 cm (1.39-6.80) and by OTO was 4.18 cm (1.63-7.09), a significant mean difference of 0.27 cm (95% CI: 0.23-0.32 cm). The reproducibility coefficients for differences between technicians were 0.30 cm (95% CI: 0.24-0.36) for ITI and 0.42 cm (95% CI: 0.35-0.49) for OTO indicating significantly better repeatability using ITI. Finally, 15 images were measured twice in random order by all screeners and sonographers. For AAAs > 5 cm, repeatability was significantly better with ITI than OTO (0.14 vs. 0.21; p = 0.016). There was the expected difference in AAA diameter between the two methods (0.27 cm). However, ITI wall method was measurably more reproducible.