Kit preparation of technetium-99m-mercaptoacetyltriglycine: analysis, biodistribution and comparison with technetium-99m-DTPA in patients with impaired renal function.

Bannister, K M; Penglis, S; Bellen, J C; Baker, R J; Chatterton, B E · J Nucl Med · 1990

case_control · Level III

Where this comes from

Abstract

Technetium-99m-mercaptoacetyltriglycine (99mTc-MAG3) was prepared by a frozen solution method, enabling the preparation of kits yielding a product substantially free of lipophilic impurities (96% 99mTc-MAG3). However, biliary activity was not completely eliminated as HPLC-purified 99mTc-MAG3 was also excreted by that route. Sequential 99mTc-DTPA and 99mTc-MAG3 renal scans were performed in 15 patients with renal dysfunction, including renal transplant recipients. In all cases, the 99mTc-MAG3 kit preparation provided superior images to 99mTc-DTPA at all levels of renal function due to a higher target-to-background ratio and a plasma clearance twice as fast as 99mTc-DTPA. Interpretation of delayed 99mTc-MAG3 images, however, was complicated by biliary excretion which will limit quantitative estimates of renal clearance. A 99mTc-MAG3 kit is likely to be of value in renal transplant assessment and in cases of significant renal impairment but would not appear to offer major advantages over 99mTc-DTPA in routine renal imaging.