Cementation of tibial inserts into nonmatching base plates in revision total knee arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21477970.
- Also identified by DOI 10.1016/j.arth.2011.02.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe the use of a novel technique for adapting nonmatching tibial inserts into tibial trays in revision total knee arthroplasty. From 1998 to 2003, the senior author performed 7 revision total knee arthroplasty procedures, during which a nonmatching tibial insert was cemented into a tibial tray to retain a well-fixed but incompatible opposite component. Bench tests were undertaken to confirm the stability of cement as a locking mechanism substitute. Three components completed 1 000 000 cycles of loading under simulated physiologic stresses with no evidence of fixation failure. There have been no clinical failures at 18 to 69 months in vivo follow-up (mean, 49 months). This technique provided durable fixation while avoiding host bone damage that might have occurred needlessly had the well-fixed implant been extracted. level IV therapeutic study, case series.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Cementation
- Knee Prosthesis
- Tibia
Anatomy
- knee
- tibia