Retinal oximetry based on nonsimultaneous image acquisition using a conventional fundus camera.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21478073.
- Also identified by DOI 10.1109/TMI.2011.2140329.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To measure the retinal arteriole and venule oxygen saturation (SO(2)) using a conventional fundus camera, retinal oximetry based on nonsimultaneous image acquisition was developed and evaluated. Two retinal images were sequentially acquired using a conventional fundus camera with two bandpass filters (568 nm: isobestic, 600 nm: nonisobestic wavelength), one after another, instead of a built-in green filter. The images were registered to compensate for the differences caused by eye movements during the image acquisition. Retinal SO(2) was measured using two wavelength oximetry. To evaluate sensitivity of the proposed method, SO(2) in the arterioles and venules before and after inhalation of 100% O(2) were compared, respectively, in 11 healthy subjects. After inhalation of 100% O(2), SO(2) increased from 96.0 ±6.0% to 98.8% ±7.1% in the arterioles (p=0.002) and from 54.0 ±8.0% to 66.7% ±7.2% in the venules (p=0.005) (paired t-test, n=11). Reproducibility of the method was 2.6% and 5.2% in the arterioles and venules, respectively (average standard deviation of five measurements, n=11).
Medical subject headings
- Diagnostic Techniques, Ophthalmological
- Fundus Oculi
- Oximetry
- Oxygen
- Retinal Vessels