Reliability and validity of the 3DNX accelerometer during mechanical and human treadmill exercise testing.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21483427.
- Also identified by DOI 10.1038/ijo.2011.39.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study aimed to assess the validity and reliability of the 3DNX triaxial accelerometer in a laboratory and mechanical setting. Experiment 1: A total of 10 units were fixed to a Multi-axis shaker table. A schedule comprising a range of accelerations was repeated on two occasions along each of the three measurement axes. Experiment 2: A total of 11 recreationally active individuals completed a treadmill protocol (4-20 km hr(-1)) on two occasions. Two 3DNX accelerometers were secured to the lower back, logging data every 5 s. Oxygen uptake was measured using the Douglas Bag technique. Reliability inter- and intra-units were assessed using a combination of limits of agreement (LoA), coefficient of variation (CV) and intra-class correlation coefficient (ICC). Validity was assessed using simple linear regression. All data are expressed in counts per 5 s (counts 5 s(-1)). Experiment 1: CV(intra) ranged from 0.0 to 8.9% in all axes. The absolute bias±95% LoA values were all below four counts 5 s(-1). CV(inter) did not rise above 4.5%. ICCs were 1.0 for all stages in all axes. The relationship between 3DNX counts and acceleration yielded an R(2) value of 0.99 and a standard error of the estimate (SEE) of six counts 5 s(-1). Experiment 2: CV(inter) ranged from 7.7 to 16.0% for trial 1 and from 7.7 to 16.2% for trial 2. ICCs between units ranged from 0.95 to 1.00 for trial 1 and from 0.90 to 0.99 for trial 2. Mean ratio bias±95% LoA values for intra-unit and inter-unit reliability were -0.7% (P>0.05)±12.4% and 0.4% (P>0.05)±5.6%, respectively. The relationship between VO(2) and 3DNX counts for walking and running was linear (R(2)=0.65, SEE=1.42 ml kg(-1) min(-1); R(2)=0.62, SEE=3.63 ml kg(-1) min(-1)). The 3DNX accelerometer is a reliable and valid device for measuring acceleration in a mechanical setting and during human treadmill exercise.
Medical subject headings
- Equipment Design
- Exercise Test
- Monitoring, Ambulatory
- Oxygen