Clinical evaluation of the i-STAT kaolin activated clotting time (ACT) test in different clinical settings in a large academic urban medical center: comparison with the Medtronic ACT Plus.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 21502428.
- Also identified by DOI 10.1309/AJCPSF8ASGONNQM6.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Historically, it has been difficult for hospitals to change methods for activated clotting time (ACT) testing because of differences in ACT values obtained with different instruments, wide differences in target ranges used in different procedures, and the difficulty of performing crossover studies at the bedside in critical care situations. There are limited published data comparing the i-STAT (Abbott Point of Care, Princeton, NJ) kaolin ACT with the Medtronic ACT Plus (Medtronic, Minneapolis, MN). The i-STAT system can perform ACT testing in addition to testing of a number of critical care analytes and may offer potential advantages over other ACT analyzers. Comparison of ACT values on 121 simultaneous split-sample tests yielded an R(2) of 0.88 with i-STAT = 0.79 Medtronic + 72.0. The Pearson correlation was R = 0.94, indicating statistically significant correlation between the 2 methods. Based on this comparison, we were able to implement the i-STAT ACT throughout our institution without changing target ranges for any individual procedure.
Medical subject headings
- Whole Blood Coagulation Time