Fracture of the outer metallic head of the bipolar hip prosthesis: an unusual bearing surface failure.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 21507603.
- Also identified by DOI 10.1016/j.arth.2011.03.018.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fracture of the bearing surface is an infrequent cause of failure of a hip arthroplasty. Although well documented with ceramic heads, fracture of the metallic head is much rarer. We report a case of a fracture of the outer metallic head of a modular cemented bipolar hemiarthroplasty 2 years after the index procedure. Over time, the outer head lost its intended motion and assumed a vertical position. We hypothesized that this position caused asymmetrical loading with stress concentration at the poles, compounded by repeated impingement between the skirted inner cobalt-chromium (Cr-Co) head and the outer stainless steel head of this particular prosthesis. These were supported by the finite element studies. In addition, scanning electron microscopy and energy dispersive x-ray studies showed metallurgical defects that seemed to have initiated and/or accelerated the fracture. Although rare, this mode of failure calls for increased awareness, periodic follow-up, and quality control.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femoral Neck Fractures
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- hip
- femur