Clinically accurate fetal ECG parameters acquired from maternal abdominal sensors.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21514560.
- Also identified by DOI 10.1016/j.ajog.2011.02.066 and PMC identifier 3145045.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We sought to evaluate the accuracy of a novel system for measuring fetal heart rate (FHR) and ST-segment changes using noninvasive electrodes on the maternal abdomen. Fetal electrocardiograms were recorded using abdominal sensors from 32 term laboring women who had a fetal scalp electrode (FSE) placed for a clinical indication. Good-quality data for FHR estimation were available in 91.2% of the FSE segments and 89.9% of the abdominal electrode segments. The root mean square error between the FHR data calculated by both methods over all processed segments was 0.36 beats per minute. ST deviation from the isoelectric point ranged from 0-14.2% of R-wave amplitude. The root mean square error between the ST change calculated by both methods averaged over all processed segments was 3.2%. FHR and ST change acquired from the maternal abdomen is highly accurate and, on average, is clinically indistinguishable from FHR and ST change calculated using FSE data.
Medical subject headings
- Electrocardiography
- Fetal Monitoring