Cobalt and chromium levels in blood and urine following hip resurfacing arthroplasty with the Conserve Plus implant.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21543699.
- Also identified by DOI 10.2106/JBJS.J.01721.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of the present study was to determine cobalt and chromium ion levels in the blood and urine of patients in whom a modern-generation metal-on-metal hip resurfacing device had been implanted. A total of ninety-seven patients with a Conserve Plus metal-on-metal hip resurfacing implant were followed prospectively for two years. Cobalt and chromium levels in erythrocytes, serum, and urine were measured preoperatively as well as three, six, twelve, and twenty-four months postoperatively. The median serum cobalt and chromium ion levels were 1.04 μg/L (range, 0.31 to 7.42 μg/L) and 2.00 μg/L (range, 0.28 to 10.49 μg/L), respectively, at one year after surgery and 1.08 μg/L (range, 0.44 to 7.13 μg/L) and 1.64 μg/L (range, 0.47 to 10.95 μg/L), respectively, at two years after surgery. The corresponding mean levels (and standard deviations) of serum cobalt and chromium were 1.68 ± 1.66 μg/L and 2.70 ± 2.22 μg/L, respectively, at one year after surgery and 1.79 ± 1.66 μg/L and 2.70 ± 2.37 μg/L, respectively, at two years after surgery. These levels compare favorably with other published ion results for metal-on-metal hip resurfacing and replacement implants. No pseudotumors or other adverse soft-tissue reactions were encountered in our study population. Further research is needed to determine the clinical importance of increased cobalt and chromium ion levels in serum and urine following metal-on-metal hip resurfacing.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Chromium
- Cobalt
- Hip Prosthesis
Anatomy
- hip