Fixation and bone remodeling around a low-modulus stem seven-year follow-up of a randomized study with use of radiostereometry and dual-energy x-ray absorptiometer.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 21549557.
- Also identified by DOI 10.1016/j.arth.2011.03.029.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Thirty-eight patients (40 hips) randomly received either an uncemented fully porous-coated composite stem (Epoch; Zimmer, Warsaw, Ind) or an uncemented proximally porous-coated solid stem (Anatomic; Zimmer). Patients were followed up for 7 years using radiostereometry, dual-energy x-ray absorptiometry, conventional radiography, the Harris Hip Score, and a pain questionnaire. Both stem designs achieved excellent outcome for fixation (stem subsidence and stem rotations close to zero) and clinical outcome, without any difference between the 2 groups (P > .12). Median wear rates were low despite use of conventionally gamma-sterilized polyethylene. No stem was radiographically loose on the postoperative radiographs. The low-modulus composite stem had positive effects on early proximal bone remodeling in Gruen regions 1, 2, 6, and 7 (P < .04). However, at 7 years, this bone-sparing effect persisted in only the calcar region (Gruen region 7).
Medical subject headings
- Absorptiometry, Photon
- Bone Remodeling
- Hip Prosthesis
- Prosthesis Retention
- Radiostereometric Analysis
Anatomy
- hip