Characterization methods of bone-implant-interfaces of bioresorbable and titanium implants by fracture mechanical means.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21565724.
- Also identified by DOI 10.1016/j.jmbbm.2010.08.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bioresorbable materials for implants have become increasingly researched over the last years. The bone-implant-interfaces of three different implant materials, namely a new bioresorbable magnesium alloy, a new self-reinforced polymer implant and a conventional titanium alloy, were tested using various methods: push-out tests, SEM and EDX analyses as well as surface analyses based on stereoscopic 3D pictures were conducted. The fracture energy is proposed as a very significant reference value for characterizing the mechanical performance of a bone-implant system. By using a video-extensometer system instead of, as is commonly done, tracking the movement of the crosshead in the push-out tests, the accuracy of measurement could be increased.