Characterization methods of bone-implant-interfaces of bioresorbable and titanium implants by fracture mechanical means.

Tschegg, E K; Lindtner, R A; Doblhoff-Dier, V; Stanzl-Tschegg, S E; Holzlechner, G; Castellani, C; Imwinkelried, T; Weinberg, A · J Mech Behav Biomed Mater · 2011

biomechanical · Level V

Where this comes from

Abstract

Bioresorbable materials for implants have become increasingly researched over the last years. The bone-implant-interfaces of three different implant materials, namely a new bioresorbable magnesium alloy, a new self-reinforced polymer implant and a conventional titanium alloy, were tested using various methods: push-out tests, SEM and EDX analyses as well as surface analyses based on stereoscopic 3D pictures were conducted. The fracture energy is proposed as a very significant reference value for characterizing the mechanical performance of a bone-implant system. By using a video-extensometer system instead of, as is commonly done, tracking the movement of the crosshead in the push-out tests, the accuracy of measurement could be increased.