Upside-down lens syndrome: ocular complications secondary to inverted implantation of the Kelman Multiflex anterior chamber intraocular lens.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21570051.
- Also identified by DOI 10.1016/j.ajo.2011.01.025 and PMC identifier 5410585.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To describe ocular complications resulting from inverted insertion of a Kelman Multiflex anterior chamber intraocular lens (ACIOL). Retrospective observational case series. We describe the outcomes of 4 cases of complicated cataract surgery in which a flexible open-loop polymethylmethacrylate (PMMA) ACIOL (Kelman Multiflex type) was placed in an inverted configuration, all of which were referred to a single institution. In all 4 cases, the placement of an inverted ACIOL resulted in ocular complications including chronic iritis, cystoid macular edema, pupil capture, iris adhesions, and corneal decompensation. Inadvertent inverted placement of the Kelman Multiflex anterior chamber intraocular lenses can lead to a constellation of severe ocular complications. When inserting such lenses, the surgeon should ensure that the lens is correctly oriented such that the optic vaults anteriorly and not posteriorly.
Medical subject headings
- Anterior Chamber
- Lens Implantation, Intraocular
- Lenses, Intraocular
- Medical Errors
- Postoperative Complications