Comparison of surface characteristics of retrieved cobalt-chromium femoral heads with and without ion implantation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21570804.
- Also identified by DOI 10.1016/j.arth.2011.03.009.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Nitrogen ion implantation of CoCr is reported to produce increased surface hardness and a lower friction surface. Femoral heads with and without ion implantation retrieved from 1997 to 2003 were evaluated for surface roughness (average surface roughness [Ra], mean peak height [Rpm], and maximum distance from peak to valley [Rmax]), nanohardness, and the ion-treated layer thickness. The difference in average Rmax (P = .033) and average Rpm (P = .008) was statistically significant, but there was no correlation between the average or maximum roughness parameters (average surface roughness, Rmax, and Rpm) and time in vivo (P > .05). Overall, nanohardness was greater for the low-friction ion-treated heads (P < .001); and it decreased with increasing time in vivo (P = .01). Ion treatment produces an increased surface hardness, but the advantage of this increased hardness appears to dissipate over time in vivo.
Medical subject headings
- Chromium
- Cobalt
- Hardness Tests
- Hip Prosthesis
- Materials Testing
Anatomy
- femur
- hip