Evaluation of the Beckman Coulter UniCel DxH 800 and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on pediatric and neonatal specimens in a tertiary care hospital.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 21571966.
- Also identified by DOI 10.1309/AJCP2EXNSLGGRVSQ.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the new UniCel DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) using 156 pediatric specimens in Microtainer tubes (Becton Dickinson, Franklin Lakes, NJ). The CBC and differential showed good interinstrument correlation, including WBCs (r = 0.995), RBCs (r = 0.992), hemoglobin (r = 0.998), mean corpuscular volume (r = 0.988), platelets (r = 0.997), neutrophils (r = 0.988), lymphocytes (r = 0.984), monocytes (r = 0.815), eosinophils (r = 0.840), basophils (r = 0.049), and nucleated RBCs (NRBCs; r = 0.906). In the instrument vs 400-cell manual differential comparison, the DxH 800 and Sapphire showed comparable performance for nearly all parameters except for NRBCs, for which the DxH 800 correlated better (r = 0.989) than the Sapphire (r = 0.906). We also compared clinical efficiency by determining whether flagged specimens showed abnormalities on a peripheral blood smear as defined by International Council for Standardization in Haematology criteria. The efficiency of the DxH 800 was 78.0% vs the Sapphire at 68.1%. Both instruments showed identical sensitivity (91.1%), but the specificity for the DxH 800 (71.9%) was higher than that of the Sapphire (57.3%).
Medical subject headings
- Blood Cell Count
- Equipment and Supplies, Hospital
- Hematology
- Hospitals