Computer-assisted femoral head-neck osteochondroplasty using a surgical milling device an in vitro accuracy study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21621956.
- Also identified by DOI 10.1016/j.arth.2011.03.048.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Surgical navigation might increase the safety of osteochondroplasty procedures in patients with femoroacetabular impingement. Feasibility and accuracy of navigation of a surgical reaming device were assessed. Three-dimensional models of 18 identical sawbone femora and 5 cadaver hips were created. Custom software was used to plan and perform repeated computer-assisted osteochondroplasty procedures using a navigated burr. Postoperative 3-dimensional models were created and compared with the preoperative models. A Bland-Altmann analysis assessing α angle and offset ratio accuracy showed even distribution along the zero line with narrow confidence intervals. No differences in α angle and offset ratio accuracy (P = 0.486 and P = 0.2) were detected between both observers. Planning and conduction of navigated osteochondroplasty using a surgical reaming device is feasible and accurate.
Medical subject headings
- Femoracetabular Impingement
- Femur Head
- Femur Neck
- Orthopedic Procedures
- Surgery, Computer-Assisted