Calculating the mounting parameters for Taylor Spatial Frame correction using computed tomography.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 21637125.
- Also identified by DOI 10.1097/BOT.0b013e3181ee40c5.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Taylor Spatial Frame uses a computer program-based six-axis deformity analysis. However, there is often a residual deformity after the initial correction, especially in deformities with a rotational component. This problem can be resolved by recalculating the parameters and inputting all new deformity and mounting parameters. However, this may necessitate repeated x-rays and delay treatment. We believe that error in the mounting parameters is the main reason for most residual deformities. To prevent these problems, we describe a new calculation technique for determining the mounting parameters that uses computed tomography. This technique is especially advantageous for deformities with a rotational component. Using this technique, exact calculation of the mounting parameters is possible and the residual deformity and number of repeated x-rays can be minimized. This new technique is an alternative method to accurately calculating the mounting parameters.
Medical subject headings
- Congenital Abnormalities
- External Fixators
- Tibia
- Tomography, X-Ray Computed