Insulation failure in robotic and laparoscopic instrumentation: a prospective evaluation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21640966.
- Also identified by DOI 10.1016/j.ajog.2011.03.055.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to detect the incidence, prevalence, and location of insulation failures (IFs) in laparoscopic and robotic instruments. In phase A, a total of 78 robotic and 298 laparoscopic instruments were tested at 20 W and 2.64 kV at Mayo Clinic in Arizona. In phase B, 60 robotic and 308 laparoscopic instruments were tested at 20 W/1 kV and 20 W/4.2 kV, respectively. In phase A, the robotic group showed a higher prevalence (25/78; 32%) and incidence of IFs after 10 uses (35/44 instruments; 80%) when compared with laparoscopy (prevalence, 39/298 [13%]; incidence, 68/189 [36%]; P<.05). In phase B, IFs were detected in 81.7% of the robotic instruments and in 19.5% of the laparoscopic instruments (P<.005). There is a high incidence and prevalence of IF in endoscopic instrumentation that is more common in the robotic group.