Cot-side electroencephalography monitoring is not clinically useful in the detection of mild neonatal hypoglycemia.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21658714.
- Also identified by DOI 10.1016/j.jpeds.2011.04.026.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine whether there is a relationship between electroencephalography patterns and hypoglycemia, by using simultaneous cot-side amplitude integrated electroencephalography (aEEG) and continuous interstitial glucose monitoring, and whether non-glucose cerebral fuels modified these patterns. Eligible babies were ≥ 32 weeks gestation, at risk for hypoglycemia, and admitted to the neonatal intensive care unit. Electrodes were placed in C3-P3, C4-P4 O1-O2 montages. A continuous interstitial glucose sensor was placed subcutaneously, and blood glucose was measured by using the glucose oxidase method. Non-glucose cerebral fuels were measured at study entry, exit, and during recognized hypoglycemia. A total of 101 babies were enrolled, with a median weight of 2179 g and gestation of 35 weeks. Twenty-four of the babies had aEEG recordings, and glucose concentrations were low (< 2.6 mM). There were 103 episodes of low glucose concentrations lasting 5 to 475 minutes, but no observable changes in aEEG variables. Plasma concentrations of lactate, beta-hydroxybutyrate, and glycerol were low and did not alter during hypoglycemia. Cot-side aEEG was not useful for the detection of neurological changes during mild hypoglycemia. Plasma concentrations of non-glucose cerebral fuels were low and unlikely to provide substantial neuroprotection.
Medical subject headings
- Blood Glucose
- Electroencephalography
- Hypoglycemia
- Monitoring, Physiologic
- Point-of-Care Systems