Comparison of radiation dose and image quality: 320-MDCT versus 64-MDCT coronary angiography.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21701026.
- Also identified by DOI 10.2214/AJR.10.5250 and PMC identifier 3294285.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to compare radiation dose and image quality of 320- and 64-MDCT angiography using prospective gating. One hundred seventy-four patients underwent 320-MDCT, and 95 patients underwent 64-MDCT. The scan parameters for 320-MDCT were 120 kVp, 400 mA, and gantry rotation of 350 milliseconds; the parameters for 64-MDCT were 120 kVp, 600 mA, and gantry rotation of 350 milliseconds. Effective dose (ED) was calculated from the dose-length product and a conversion factor (k = 0.014 mSv / mGy × cm). Two observers independently assessed image quality using a 3-point scale, where 3 denotes excellent quality and 1 denotes nondiagnostic quality, using a 16-segment model. Discrepancies were settled by consensus. The ED was significantly lower in patients undergoing 320-MDCT angiography, with a median ED of 4.4 mSv (interquartile range [IQR], 3.4-6.2 mSv), compared with 64-MDCT angiography, with a median ED of 6.2 mSv (IQR, 5.5-6.9 mSv) (p = 0.0001). In patients with a heart rate of 65 beats/min or less (92%), the median radiation dose using 320-MDCT was 4.1 mSv (IQR, 3.2-6.1 mSv), and that for 64-MDCT angiography was 6.2 mSv (IQR, 5.8-6.9 mSv) (p = 0.0001). In patients with heart rate greater than 65 beats/min (8%), the median dose was higher with 320-MDCT (8.7 mSv; IQR, 5.9-14.3 mSv) than with 64-MDCT (5.8 mSv; IQR, 5.3-6.7 mSv) (p = 0.02). Segmental image quality was significantly better for 320-MDCT (excellent or good quality, 96.66%; nondiagnostic quality, 0.1%) than for 64-MDCT angiography (excellent or good quality, 86%; nondiagnostic quality, 3.33%) (all p < 0.0001). Image quality was good for both 320- and 64-MDCT angiography. Overall radiation dose was significantly lower in 320-MDCT angiography when the heart rate was 65 beats/min or less. Every effort should be made to control heart rate to minimize radiation dose.
Medical subject headings
- Coronary Angiography
- Coronary Artery Disease
- Radiation Dosage
- Tomography, X-Ray Computed