Valve-in-valve hemodynamics of 20-mm transcatheter aortic valves in small bioprostheses.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21704287.
- Also identified by DOI 10.1016/j.athoracsur.2011.04.009.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Transcatheter aortic valve (TAV) implantation is a treatment for selected patients with failing bioprostheses. We previously showed that currently available SAPIEN (Edwards Lifesciences, Irvine, CA) TAV sizes did not yield acceptable valve-in-valve (VIV) hemodynamics in small degenerated bioprostheses because optimal TAV function requires full stent expansion to its nominal size. The study objective was to determine (1) if 20-mm TAVs provide acceptable hemodynamics in small degenerated bioprostheses and (2) the effect of TAV spatial orientation on valvular hemodynamics and coronary flows. Twelve 20-mm TAVs were created for implantation within 19-mm and 21-mm degenerated Carpentier-Edwards Perimount (Edwards Lifesciences) and porcine bioprostheses. Degenerated valves were sutured into human homograft roots and mounted in a pulse duplicator. TAVs were implanted within bioprostheses as VIV in standard orientation, in which TAV and bioprosthetic commissures were aligned, and later with 60-degree rotation. The 20-mm TAVs migrated retrograde into the left ventricle after VIV in the 21-mm Perimount bioprostheses. However, 20-mm TAVs in 19-mm Perimount (54.9±5.4 to 23.5±3.9 mm Hg, p=0.006) and 21-mm porcine bioprostheses (35.2±8.9 to 16.8±4.1 mm Hg, p=0.03) significantly reduced mean gradients. No significant reduction in pressure gradient occurred after VIV in 19-mm degenerated porcine bioprostheses. Mild regurgitation was observed after VIV. VIV with standard and 60-degree TAV orientation did not significantly alter hemodynamics or coronary flows. Valve-in-valve hemodynamics with 20-mm TAV improved for 19-mm Perimount and 21-mm porcine but not 19-mm porcine bioprostheses. No significant differences in hemodynamics were noted by orientation with TAV and bioprosthesis commissural alignment or 60-degree rotation.
Medical subject headings
- Aortic Valve
- Aortic Valve Stenosis
- Bioprosthesis
- Cardiac Catheterization
- Heart Valve Prosthesis Implantation
- Hemodynamics
- Prosthesis Design
- Prosthesis Failure
- Prosthesis Fitting