Facile fabrication of a rigid and chemically resistant micromixer system from photocurable inorganic polymer by static liquid photolithography (SLP).
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21713287.
- Also identified by DOI 10.1039/c1lc20118b.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Highly effective mixing in microchannels is important for most chemical reactions conducted in microfluidic chips. To obtain a rigid and chemically resistant micromixer system at low cost, we fabricated a Y-shaped microchannel with built-in mixer structures by static liquid photolithography (SLP) from methacrylated polyvinylsilazane (MPVSZ) as an inorganic polymer photoresist which was then converted to a silicate phase by hydrolysis in vaporized ammonia atmosphere at 80 °C. The microchannel incorporating herringbone mixer structures was bonded with a matching polydimethylsiloxane (PDMS) open channel which was pre-coated by perhydropolysilazane (PHPS)-based mixture, and finally treated by additional hydrolysis at room temperature to convert the PHPS layer to a silica phase. Finally, the chemical resistance of the microfluidic system with embedded micromixer was confirmed with various solvents, and the excellent mixing performance in a short mixing length of 2.3 cm was demonstrated by injecting two different colored fluids into the microchannel.
Medical subject headings
- Dimethylpolysiloxanes
- Microfluidic Analytical Techniques
- Polymers