Interface micromotion of uncemented femoral components from postmortem retrieved total hip replacements.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21723696.
- Also identified by DOI 10.1016/j.arth.2011.04.018 and PMC identifier 3202685.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Axial torsional loads representative of gait and stair climbing conditions were applied to transverse sections of 8 uncemented postmortem retrievals and a high-resolution imaging system with digital image correlation was used to measure local micromotion along the bone-implant interface. For 7 components that were radiographically stable, there was limited micromotion for gait loading (1.42 ± 1.33 μm) that increased significantly (P = .0032) for stair climb loading (7.32 ± 9.96 μm). A radiographically loose component had motions on the order of 2.3 mm with gait loading. There was a strong inverse relationship between the amount of bone-implant contact (contact fraction) (P = .001) and micromotion. The uncemented components had greater contact fraction (41.8% ± 14.4% vs 11.5% ± 10.2%, P = .0033) and less median micromotion (0.81 ± 0.79 μm vs 28.8 ± 51.1 μm) compared to a previously reported study of cemented retrievals.