Comparison of manually corrected retinal thickness measurements from multiple spectral-domain optical coherence tomography instruments.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21733920.
- Also identified by DOI 10.1136/bjo.2010.201111.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare retinal thickness measurements from three different spectral domain optical coherence instruments when manual segmentation is employed to standardise retinal boundary locations. 40 eyes of 21 healthy subjects were scanned on the Cirrus HD-OCT, Topcon 3D-OCT-2000 and Heidelberg Spectralis-OCT. Raw data were imported into custom grading software (3D-OCTOR). Manual segmentation was performed on every data set, and retinal thickness values in the foveal central subfield were computed. 37 eyes of 20 subjects were gradable on every machine. The average retinal thicknesses for these eyes were 236.7 μm (SD 20.1), 235.7 μm (SD 20.4) and 236.5 μm (SD 18.0) for the Cirrus, 3D-OCT-2000 and Spectralis, respectively. Comparing manual retinal thickness measurements between any two machines, the maximum difference was 18.2 μm. The mean absolute differences per eye between two machines were: 4.9 μm for Cirrus versus 3D-OCT-2000, 3.7 μm for Cirrus versus Spectralis and 4.4 μm for 3D-OCT-2000 versus Spectralis. When a uniform position is used to locate the outer retinal boundary, the retinal thickness measurements derived from three different spectral domain optical coherence instruments devices are virtually identical. Manual correction may allow OCT-derived thickness measurements to be compared between devices in clinical trials and clinical research.
Medical subject headings
- Diagnostic Techniques, Ophthalmological
- Retina
- Tomography, Optical Coherence