Comparison of in vitro hydrolysis, subcutaneous and intramedullary implantation to evaluate the strength retention of absorbable osteosynthesis implants.

Vasenius, J; Vainionpää, S; Vihtonen, K; Mäkelä, A; Rokkanen, P; Mero, M; Törmälä, P · Biomaterials · 1990

basic_science · Level V

Where this comes from

Abstract

To evaluate the behaviour of mechanical properties of absorbable osteosynthesis implants in vivo, the strength retention of self-reinforced polyglycolide rods in distilled water at 37 degrees C in the subcutis and femoral medullary cavity of the rabbit was investigated. The self-reinforced polyglycolide rods lost their strength significantly faster in vivo than in vitro. The strength loss of the self-reinforced polyglycolide rods was only slightly faster in the medullary cavity than in the subcutis. As the removal of the implants from the medullary cavity became difficult 5-6 wk after implantation, it is suggested that subcutaneous implantation would be a suitable method to evaluate the strength retention of absorbable osteosynthesis implants.