Comparison of in vitro hydrolysis, subcutaneous and intramedullary implantation to evaluate the strength retention of absorbable osteosynthesis implants.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2173632.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the behaviour of mechanical properties of absorbable osteosynthesis implants in vivo, the strength retention of self-reinforced polyglycolide rods in distilled water at 37 degrees C in the subcutis and femoral medullary cavity of the rabbit was investigated. The self-reinforced polyglycolide rods lost their strength significantly faster in vivo than in vitro. The strength loss of the self-reinforced polyglycolide rods was only slightly faster in the medullary cavity than in the subcutis. As the removal of the implants from the medullary cavity became difficult 5-6 wk after implantation, it is suggested that subcutaneous implantation would be a suitable method to evaluate the strength retention of absorbable osteosynthesis implants.
Medical subject headings
- Fracture Fixation, Intramedullary
- Materials Testing
- Prostheses and Implants