Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography.
Level III
Where this comes from
- Record sourced from PubMed, PMID 21761267.
- Also identified by DOI 10.1007/s00259-011-1877-y.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the diagnostic accuracy of attenuation corrected nuclear myocardial perfusion imaging (MPI) with a novel hybrid single photon emission computed tomography (SPECT)/CT device consisting of an ultrafast dedicated cardiac gamma camera with cadmium-zinc-telluride (CZT) solid-state semiconductor detectors integrated onto a multislice CT scanner to detect coronary artery disease (CAD). Invasive coronary angiography served as the standard of reference. The study population included 66 patients (79% men; mean age 63 ± 11 years) who underwent 1-day (99m)Tc-tetrofosmin pharmacological stress/rest examination and angiography within 3 months. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) as well as accuracy of the CT X-ray based attenuation corrected CZT MPI for detection of CAD (≥ 50% luminal narrowing) was calculated on a per-patient basis. The prevalence of angiographic CAD in the study population was 82%. Sensitivity, specificity, PPV, NPV and accuracy were 87, 67, 92, 53 and 83%, respectively. In this first report on CZT SPECT/CT MPI comparison versus angiography we confirm a high accuracy for detection of angiographically documented CAD.
Medical subject headings
- Cadmium
- Coronary Angiography
- Multimodal Imaging
- Myocardial Perfusion Imaging
- Positron-Emission Tomography
- Tellurium
- Tomography, X-Ray Computed
- Zinc