Long term in vitro functional stability and recording longevity of fully integrated wireless neural interfaces based on the Utah Slant Electrode Array.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21775785.
- Also identified by DOI 10.1088/1741-2560/8/4/045004 and PMC identifier 3235955.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluate the encapsulation and packaging reliability of a fully integrated wireless neural interface based on a Utah Slant Electrode Array/integrated neural interface-recording version 5 (USEA/INI-R5) system by monitoring the long term in vitro functional stability and recording longevity. The INI encapsulated with 6 µm Parylene-C was immersed in phosphate buffered saline (PBS) for a period of over 276 days (with the monitoring of the functional device still ongoing). The full functionality (wireless radio-frequency power, command and signal transmission) and the ability of the electrodes to record artificial neural signals even after 276 days of PBS soaking with little change (within 14%) in signal/noise amplitude constitute a major milestone in long term stability and allow us to study and evaluate the encapsulation reliability, functional stability and its potential usefulness for a wireless neural interface for future chronic implants.
Medical subject headings
- Electrodes, Implanted
- Prosthesis Design
- User-Computer Interface
- Wireless Technology