Failure analysis of the fractured wires in sternal perichronal loops.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 21783110.
- Also identified by DOI 10.1016/j.jmbbm.2011.02.015.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report failure analysis of sternal wires in two cases in which a perichronal fixation technique was used to close the sternotomy. Various characteristics of the retrieved wires were compared to those of unused wires of the same grade and same manufacturer and with surgical wire specifications. In both cases, wire fracture was un-branched and transgranular and proceeded by a high cycle fatigue process, apparently in the absence of corrosion. However, stress anlysis indicates that the effective stress produced during strong coughing is lower than the yield strength. Our findings suggest that in order to reduce the risk for sternal dehiscence, the diameter of the wire used should be increased.
Medical subject headings
- Equipment Failure Analysis
- Mechanical Phenomena
- Sternotomy
- Suture Techniques