Impact of thermomechanical texture on the superelastic response of Nitinol implants.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21783153.
- Also identified by DOI 10.1016/j.jmbbm.2011.05.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The phenomenon of superelasticity in near-equiatomic NiTi, which originates from a first-order martensitic phase transition, is exploited in an increasing number of biomedical devices, most importantly endovascular stents. These stents are often manufactured from microtubing, which is shown to be highly textured crystallographically. Synchrotron X-ray microdiffraction provided microstructural, phase, and strain analysis from Nitinol tube sections that were deformed in situ along longitudinal, circumferential, and transverse orientations. We show that the large variation in the superelastic response of NiTi in these three tube directions is strongly influenced by the path that the martensitic transformation follows through the microstructure. Specifically, in severely worked NiTi, bands of [100] grains occur whose orientation deviates markedly from the surrounding matrix; these bands have an unusually large impact on the initiation and the propagation of martensite, and hence on the mechanical response. Understanding the impact of these local microstructural effects on global mechanical response, as shown here, leads to a much fuller understanding of the causes of deviation of the mechanical response from predictions and unforeseen fracture in NiTi biomedical devices.
Medical subject headings
- Alloys
- Elasticity
- Prostheses and Implants
- Temperature