Fracture of the modular femoral neck component in total hip arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 21798696.
- Also identified by DOI 10.1016/j.arth.2011.05.024.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The use of modularity, specifically dual modular femoral stems, in total hip arthroplasty has increased in popularity over the past 2 decades. While offering several distinct advantages intraoperatively, the long-term success of adding a second modular junction has yet to be established. One potential complication of increasing modularity is component fracture. We present a case of modular femoral neck prosthesis fracture necessitating revision surgery to treat this complication. Careful preoperative planning during revision of these failures is essential to avoid morbidity and unnecessary subsequent revision surgeries, as demonstrated in this case. The combined effects of crevice and fretting corrosion, large-diameter femoral head, long modular neck, metal-on-metal articulation, patient size, and activity level may have all played integral roles in creating an environment susceptible to this classic pattern of fatigue fracture.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur Head Necrosis
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- femur
- hip