Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 21822785.
- Also identified by DOI 10.1007/s00330-011-2227-y.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate prospectively, in patients undergoing body CTA, the radiation dose saving potential of raw data-based iterative reconstruction as compared to filtered back projection (FBP). Twenty-five patients underwent thoraco-abdominal CTA with 128-slice dual-source CT, operating both tubes at 120 kV. Full-dose (FD) images were reconstructed with FBP and were compared to half-dose (HD) images with FBP and HD-images with sinogram-affirmed iterative reconstruction (SAFIRE), both reconstructed using data from only one tube-detector-system. Image quality and sharpness of the aortic contour were assessed. Vessel attenuation and noise were measured, contrast-to-noise-ratio was calculated. Noise as image quality deteriorating artefact occurred in 24/25 (96%) HD-FBP but not in FD-FBP and HD-raw data-based iterative reconstruction datasets (p < 0.001). Other artefacts occurred with similar prevalence among the datasets. Sharpness of the aortic contour was higher for FD-FBP and HD-raw data-based iterative reconstruction as compared to HD-FBP (p < 0.001). Aortoiliac attenuation was similar among all datasets (p > 0.05). Lowest noise was found for HD-raw data-based iterative reconstruction (7.23HU), being 9.4% lower than that in FD-FBP (7.98HU, p < 0.05) and 30.8% lower than in HD-FBP images (10.44HU, p < 0.001). Contrast-to-noise-ratio was lower in HD-FBP (p < 0.001) and higher in HD-raw data-based iterative reconstruction (p < 0.001) as compared to FD-FBP. Intra-individual comparisons of image quality of body CTA suggest that raw data-based iterative reconstruction allows for dose reduction >50% while maintaining image quality. Key Points • Raw data-based iterative reconstruction reduces image noise and improves image quality as compared to filtered back projection • At a similar radiation dose, raw data-based iterative reconstruction improves the sharpness of vessel contours • In body CTA a dose reduction of >50% might be possible when using raw data-based iterative reconstructions, while image quality can be maintained.
Medical subject headings
- Aorta
- Contrast Media
- Coronary Angiography
- Iliac Artery
- Radiographic Image Interpretation, Computer-Assisted
- Tomography, X-Ray Computed