Endotracheal temperature and humidity in laryngectomized patients in a warm and dry environment and the effect of a heat and moisture exchanger.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 21837698.
- Also identified by DOI 10.1002/hed.21597.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study was designed to assess endotracheal climate in laryngectomized patients in a warm and dry environment and the effects of a heat and moisture exchanger (HME). Endotracheal temperature and humidity were measured in 11 laryngectomized patients with a regularly used HME (Provox Normal HME; R-HME), an HME with antimicrobial filter (Provox Micron HME; F-HME), and without HME (open stoma). Measurements were performed at 32°C, 34°C, and 38°C (<25% relative humidity). Both R-HME and F-HME increased end-inspiratory humidity (AH(insp)) equally (range, 3.4 to 5.6 mg H(2)O/L). The R-HME has a cooling effect on end-inspiratory temperature (T(insp)), which is similar for all tested environmental conditions (-3.8°C); F-HME decreases T(insp) less (range, -1.3 to -0.6°C). In a warm and dry environment, both R- and F-HME significantly cool and humidify inspired air. Therefore, consistent use of an HME under these climate conditions is also probably clinically beneficial.
Medical subject headings
- Body Temperature
- Humidity
- Laryngectomy
- Respiratory Therapy
- Temperature