Metal-induced artifacts in MRI.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 21862795.
- Also identified by DOI 10.2214/AJR.11.7364 and PMC identifier 5562503.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this article is to review some of the basic principles of imaging and how metal-induced susceptibility artifacts originate in MR images. We will describe common ways to reduce or modify artifacts using readily available imaging techniques, and we will discuss some advanced methods to correct readout-direction and slice-direction artifacts. The presence of metallic implants in MRI can cause substantial image artifacts, including signal loss, failure of fat suppression, geometric distortion, and bright pile-up artifacts. These cause large resonant frequency changes and failure of many MRI mechanisms. Careful parameter and pulse sequence selections can avoid or reduce artifacts, although more advanced imaging methods offer further imaging improvements.
Medical subject headings
- Artifacts
- Magnetic Resonance Imaging
- Metals
- Prostheses and Implants