Implantation-related complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices: a systematic review of randomized clinical trials.
systematic_review · Level I
Where this comes from
- Record sourced from PubMed, PMID 21867832.
- Also identified by DOI 10.1016/j.jacc.2011.06.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The number of implantable cardioverter-defibrillator (ICD) and cardiac resynchronization therapy (CRT) implantations is increasing drastically worldwide, and hence, the number of implanting centers is also increasing. Despite abundant data on the beneficial effect of these devices, little is known regarding safety and complication rates. Eleven ICD and 7 CRT trials were systematically reviewed to provide data on the frequency of in-hospital mortality and complications related to the implantation. Average in-hospital mortality was 2.7% in trials using both thoracotomy and nonthoracotomy ICDs, 0.2% in trials using nonthoracotomy ICDs, and 0.3% in CRT trials. The pneumothorax rate was similar between the nonthoracotomy ICD and CRT trials (0.9%) Coronary sinus complications occurred in 2.0% of patients undergoing CRT. Lead dislodgement rates were higher in CRT trials (5.7%) than in nonthoracotomy ICD trials (1.8%).
Medical subject headings
- Cardiac Resynchronization Therapy Devices
- Defibrillators, Implantable
- Prosthesis Implantation