Monitoring tumor motion by real time 2D/3D registration during radiotherapy.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 21885144.
- Also identified by DOI 10.1016/j.radonc.2011.07.031 and PMC identifier 3276833.
- Licence recorded as CC BY-NC-ND.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this paper, we investigate the possibility to use X-ray based real time 2D/3D registration for non-invasive tumor motion monitoring during radiotherapy. The 2D/3D registration scheme is implemented using general purpose computation on graphics hardware (GPGPU) programming techniques and several algorithmic refinements in the registration process. Validation is conducted off-line using a phantom and five clinical patient data sets. The registration is performed on a region of interest (ROI) centered around the planned target volume (PTV). The phantom motion is measured with an rms error of 2.56 mm. For the patient data sets, a sinusoidal movement that clearly correlates to the breathing cycle is shown. Videos show a good match between X-ray and digitally reconstructed radiographs (DRR) displacement. Mean registration time is 0.5 s. We have demonstrated that real-time organ motion monitoring using image based markerless registration is feasible.
Medical subject headings
- Carcinoma, Non-Small-Cell Lung
- Lung Neoplasms
- Radiographic Image Interpretation, Computer-Assisted
- Radiotherapy Planning, Computer-Assisted
- Tomography, X-Ray Computed