A new glucose-clamp algorithm--theoretical considerations and computer simulations.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2188902.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The commonly used setup of an automatic glucose-clamp was analyzed as a control system, modeling the measuring device and the patient in a simple but effective way. Strict limits in response time, parasitic oscillation amplitude, and accuracy were defined in order to approach the physician's requirements. We developed a new control algorithm and defined the gain coefficients which can lead the system within these limits. Computer simulations with model parameters from literature (patient) and from experimental data (measuring device) are presented. Preliminary in vivo trials are also presented.
Medical subject headings
- Algorithms
- Computer Simulation
- Glucose Clamp Technique
- Models, Biological