Realisation of the absorbed dose to water for I-125 interstitial brachytherapy sources.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 21924788.
- Also identified by DOI 10.1016/j.radonc.2011.08.025.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A large air-filled parallel-plate extrapolation chamber in a phantom of water-equivalent material is used as a primary standard measuring device for low-energy interstitial brachytherapy sources from which the unit of absorbed dose to water can be derived. The chamber is suitable for low-energy photons with energies up to 50 keV. The method to determine the absorbed dose to water was newly developed based on radiation transport theory. It offers a clear analytical expression to determine D(w). A conversion factor C(x(i),x(i)(+1)) has to be applied to the difference of ionization charges measured at two plate separations x(i) and x(i)(+1). The details of the method are presented. The determination of D(w) of an I-125 seed is demonstrated by the measurement of a 'BEBIG Symmetra I25.S16' - seed. Additional measurements of the reference air kerma rate with the PTB primary standard chamber GROVEX I allow to determine experimentally the dose rate constant of an I-125 seed by means of primary standards for the first time. Good agreement was found between the obtained dose rate constant and the published data.