Methods for measuring GFR with technetium-99m-DTPA: an analysis of several common methods.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 2194005.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Several commonly used scintigraphic methods of GFR measurement were evaluated. Forty-three adult patients with a wide range of ages and renal function were studied. The two-sample plasma method of Russell and the urinary method of Jackson were the most accurate methods overall. The one-sample plasma method of Russell, the volume of distribution method of Fawdry, and a terminal slope method were less reliable, especially at low (0-60 ml/min) GFRs. The renal uptake method of Gates correlated poorly to the standards at all GFR levels even when corrected for body surface area or blood volume. The Russell two point and Jackson urinary GFR's can be used as complementary techniques and are recommended as primary methods of scintigraphic GFR determination.
Medical subject headings
- Glomerular Filtration Rate
- Organotechnetium Compounds
- Pentetic Acid